[1]孙思源.卡光杆综合测试仪校准标准化探讨[J].石油管材与仪器,2018,4(05):71-74.[doi:10.19459/j.cnki.61-1500/te.2018.05.021]
             SUN Siyuan.Calibration Standardization of Card Polished Rod Comprehensive Test Instrument[J].Petroleum Tubular Goods & Instruments,2018,4(05):71-74.[doi:10.19459/j.cnki.61-1500/te.2018.05.021]
            点击复制

            卡光杆综合测试仪校准标准化探讨
            分享到:

            《石油管材与仪器》[ISSN:2096-0077/CN:61-1500/TE]

            卷:
            4
            期数:
            2018年05期
            页码:
            71-74
            栏目:
            标准化
            出版日期:
            2018-10-20

            文章信息/Info

            Title:
            Calibration Standardization of Card Polished Rod Comprehensive Test Instrument
            文章编号:
            2096-0077(2018)05-0071-04
            作者:
            孙思源
            大庆油田有限责任公司测试技术服务分公司 黑龙江 大庆 163513
            Author(s):
            SUN Siyuan
            Logging and Testing Service Company, Daqing Oilfiled Co. Ltd., Daqing, Heilongjiang 163513,China
            关键词:
            校准标准化、载荷、卡光杆、拉伸、压载
            Keywords:
            calibration standardization load card polished rod stretching ballast
            分类号:
            P631.4+3
            DOI:
            10.19459/j.cnki.61-1500/te.2018.05.021
            文献标志码:
            A
            摘要:
            为了实现卡光杆综合测试仪量值的有效传递,分析常用几种卡光杆综合测试仪的校准项目、校准范围、校准流程、校准方法,分析结果表明卡光杆加速度综合测试仪校准标准存在问题:载荷零点存在初值,需在零点偏移、零点漂移、载荷示值校准时需要零点修正;加速?#20219;?#31227;校准标准应规定校准需要完成步骤(标定、检定)及设点、循环次数等,因此卡光杆综合测试仪校准必须完善目前的校准标准,实现标准化操作。
            Abstract:
            In order to achieve the polished rod card comprehensive test instrument value transferring effectively, the calibration program, range of the calibration, calibration process, and calibration method of several frequently used card polished rod comprehensive test instruments are analyzed. The results show that there are some problems in the acceleration card polished rod comprehensive test instrument calibration standard: initial value of load zero, zero shift, zero drift, load value calibration to zero point correction. The acceleration displacement calibration standards should stipulate the operation steps (calibration, verification) and select a spot, cycle times, etc., so the card polished rod comprehensive test instrument calibration standards must be improved to realize the standardized operation.

            参考文献/References:

            [1] 牟海维,王永涛,成国库.油井抽油机示功仪的研制[J]. 电测与仪表,2000,37(5):35-38.
            [2] ?#24179;?#19996;,吴利文,刘松林.无线监测示功仪的研制与应用[J]. 石油机械,2003,34(9):63-65.
            [3] 国家发展和改革委?#34987;?电子示功仪校准方法:SY/T 6678-2007[S].?#26412;?#30707;油工业出版社,2008:1-6.
            [4] 国家能源局.抽油机井综合测试仪校准方法:SY/T 7080-2016[S].?#26412;?#30707;油工业出版社,2016:1-5.
            [5] 国家质量监督检验检疫总局.通用计量术语及定义技术规范:JJF1001-2011[S].?#26412;?#20013;国质检出版社,2011:12.